Specifications

  • Construction = 18/19 swg
  • Triple Stage filtration
  • EU6 – Prefilter (95% down to 5 micron)
  • Hygieno brand pre filter
  • EU14 – HEPA Filter – Supply (99.999% down to 0.3 micron)
  • AAF brand pre filter
  • EU7 – INTERMEDIATE FILTER – Supply / Exhaust (95% down to 3 micron)
  • Hygieno brand pre filter
  • Balanced Motor
  • Internal covings for easy cleaning 2 side
  • SS 304 Double walled Side Panels
  • DOP Test port at upstream of HEPA filter
  • Differential Pressure Gauge : 2 Nos to monitor pressure drop across two stages of filters.
  • On / Off Switch for Motor
  • 5/15 Amp Switch Socket for external equipment
  • Fluorescent Light with Milky White Diffuser
Reverse Laminar Air Flow

Accessories Made Available

  • Alarm for blower tripping with Fault Acknowledgement & Reset Switch
  • Flame Proof Execution
  • Front Anti Static PVC Curtains (for warehouse applications)

Documents

  • MOC test certificates & OQ , PQ , IQ , DQ

Operating Conditions

  • Air cleanliness
  • ISO Class 5 (ISO 14644-1:1999 (E))
  • Air Velocities
  • 0.45 ± 0.05 mps
  • Air Flow
  • Vertical – Recirculatory
  • Noise Level
  • Less than 67 dB
  • Vibration Level
  • Minimum
  • Power Supply
  • 230V AC 1-Ø 50HZ